3 technical data, Technical data, Instrument components and specifications – Leica Biosystems RM2235 User Manual
Page 14
![background image](/manuals/565488/14/background.png)
12
Instructions for Use V 1.8 RevC – 09/2013
3.3 Technical Data
General
Approvals:
The instrument-specific marks are located on the
rear panel of the instrument next to the nameplate.
Operating temperature range:
+10 °C to +35 °C
Temperature range during storage:
+5 °C to +55 °C
Relative humidity:
max. 80 % non-condensing
Storage humidity:
< 80 %
Section thickness range:
1.0 - 60.0 µm
Section thickness settings:
from 1.0 - 10.0 µm in 1.0 µm increments
from 10.0 - 20.0 µm in 2.0 µm increments
from 20.0 - 60.0 µm in 5.0 µm increments
Specimen feed:
approx. 24 mm, ± 2 mm
Vertical stroke:
70 mm
Max. cutting range without retraction:
69 mm (without specimen orientation 1 µm)
Maximum sectioning area with retraction: 62 mm
The specimen retraction can be turned
off manually:
approx. 40 µm
Dimensions and weight
Width (including handwheel):
413 mm
Depth (including waste tray):
618 mm
Height (total):
305 mm (with tray on the hood)
Working height (knife blade):
168 mm (measured from the table)
Weight (without accessories):
approx. 37 kg
Optional equipment and optional accessories
Specimen orientation (option)
Horizontal:
8°
Vertical: 8°
Angle of rotation:
± 90°
Trimming stages:
10 µm, 50 µm
Repositioning of knife holder base
North-south:
± 25 mm
3.
Instrument components and specifications