beautypg.com

5 specimen holder with precision orientation, Operation – Leica Biosystems RM2235 User Manual

Page 23

background image

21

Leica RM2235

5. Operation

5.1.5 Specimen holder with precision orientation

Orienting the specimen

Fig. 17

Display of the zero position
For better display of the zero position, the orien-

tation has two red indicators (32).
When both indicators are visible and both set-

screws are in zero position at the same time

(notch point, white marking on ""), the specimen

is in zero position.

The specimen orientation allows for simple posi-

tion correction of the specimen surface when the

specimen is clamped into place.

32

30

32

31

29

When the large standard specimen

clamp (50 x 55 mm) is used, the speci-

men orientation of 8° in north-south

direction is no longer possible.

The usable angle is only about 4° in

this case.

Specimen blocks must NOT be orient-

ed during the retraction phase!

If a block is oriented during retraction,

the block will advance by the retrac-

tion value PLUS the selected section

thickness before the next section.

This may cause damage to both speci-

men and knife!

• Raise the specimen head to the upper end

position and activate the handwheel lock.

• To release the clamp, turn the eccentric lever

(29) forwards.

• Turn setscrew (30) to orient the specimen in

north-south direction. Turn setscrew (31) to

orient the specimen in east-west direction.

Each complete turn of the screw inclines the

specimen by 2°. A total of 4 complete turns = 8°

are possible in every direction. The accuracy

is approximately ± 0.5°.

For ease of estimation, there is a white mark on

the handle and a click stop that is noticeable

during turning.

• To lock the current orientation, turn the ec-

centric lever (29) backwards.

In the quick clamping device of the specimen holder fixture with precision orientation, all

specimen clamps available as optional accessories can be used (implemented).