beautypg.com

4 assembling the handwheel, Assembling the handwheel, Instrument setup – Leica Biosystems RM2235 User Manual

Page 18

background image

16

Instructions for Use V 1.8 RevC – 09/2013

4.

Instrument Setup

The feather key (4) is loosely

placed in the handwheel shaft (1)

and fixed in place with a cable tie

during transport.

• Remove the cable tie (3).

Caution!

Make sure not to lose the

feather key!

• Place the handwheel (2) on the

handwheel shaft (1) as shown.

• Tighten the screw (2a) located

in the center hole of the hand-

wheel with a Allen key No. 4 (5).

• Remove the cover foil from the

self-adhesive cover disk and

fix the cover disk on the hand-

wheel.

4.4 Assembling the handwheel

Fig. 9

1

3

2

2a

5

4

The handwheel has to be assembled before attempting

to use the instrument.

The necessary parts and tools can be found in the toolkit

supplied in the delivery.