4 mechanical trimming function, 3 specimen retraction, Operation – Leica Biosystems RM2235 User Manual
Page 22
![background image](/manuals/565488/22/background.png)
20
Instructions for Use V 1.8 RevC – 09/2013
5. Operation
5.1.3 Specimen retraction
The specimen retraction serves for protecting
the knife and the specimen. When the retrac-
tion is switched on, the specimen is drawn
back 40 µm into the upper end position after the
sectioning stroke during the return movement.
Before the feed motion of the new section thick-
ness, the feed motion for the retraction value
takes place.
The specimen retraction can be switched off
manually at the back of the instrument (Fig. 15), if
required, using the slotted-screwdriver supplied
with the delivery.
Before switching the specimen retraction On
and Off, run the specimen head to the upper end
position by turning the handwheel.
5.1.4 Mechanical trimming function
The RM2235 is fitted with a mechanical trimming function. The trim-
ming lever has 3 notching positions (0, 10 µm, 30 µm).
The points (36) mark the two trimming stages:
= 10 µm
= 30 µm
• For activating the trimming function, press the lever downwards
into one of the two notching positions and keep depressed. After
each rotation of the handwheel, a feed motion of 10 µm or 30 µm
takes place.
• After letting go of the lever, it automatically springs back to its
original position (zero position). The trimming function is thereby
deactivated.
Magnet for
holding the
wrench
size 4
For turning the retraction
Off, turn the slot into the
horizontal position (off).
For turning the retraction
On, turn the slot into the
vertical position (on).
Fig. 15
Fig. 16
36
The section thickness that has been set is not added to the
selected trimming value.
If the section thickness that has been set is greater than the
selected trimming value, the section thickness is fed.