2.3 Integrated safety devices
Automatic circuit breaker in the main switch
• An automatic circuit breaker with a lit
switch is located in the main switch. This
circuit breaker disconnects the unit from the
power supply in the event of a short circuit.
• In this case, the main switch jumps to the
position "0" = Off.
Fig. 2
Short-circuit
2. Safety
• The minimum voltage (see Technical data) must be maintained while starting the refrigera-
tion unit.
The compressor needs a start-up current of approx. 25
A. A stable power supply in accor-
dance with the instrument's specifications is essential to its proper functioning. Please
ensure that your electrical installation fulfills these preconditions prior to installing the unit.
Nonobservance causes damage to the instrument.
• Switch off the instrument each time before servicing, repairing or cleaning, and pull out the
power plug.
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)