3 funktionsstörungen, Instrument malfunctions, Operation – Leica Biosystems EG1150 C User Manual
Page 17

17
Leica EG1150 C
5. Operation
5.3 Instrument malfunctions
If you cannot solve your problem using the help in the following table, please contact your Leica
customer service organization or the Leica dealer from whom you purchased the instrument.
Malfunction
Possible causes
Corrective action
Instrument doesn't work.
Power switch not turned on.
or
The cut-out function of the
power switch activated.
Check instrument is connected,
then turn main power switch
on.
Only for Great Britain:
Check the fuse in the power
plug.
Cooling unsatisfactory,
formation of condensation.
Room temperature too high or
exposure to direct sunlight.
Check the cooling again at a
lower room temperature.
Inadequate air supply to
ventilation unit.
Check distance of instrument to
wall; it must be at least 15 cm.
Refrigeration system leak.
The instrument must be
serviced or repaired by
a service technician.
Please contact customer
service.
Compressor switches off after
a brief period of operation
and switches back on shortly
thereafter.
AC power does not match
specifications or
Compressor defective.
Check the voltage of power
supply.
Compressor does not start.
Compressor defective.
The instrument must be
serviced or repaired by
a service technician.
Please contact customer
service.
Compressor switches off after
a short period of operation.
Compressor defective.