Instrument components and specifications – Leica Biosystems EG1150 C User Manual
Page 11

11
Leica EG1150 C
3.2 Technical data
General data
Approvals:
The instrument-specific marks are located on the rear
panel of the instrument next to the name plate
The 120 V version has the test symbol c-CSA-us.
Voltage of power supply:
The device requires a specific AC volt-
age, see table at right.
Instrument type acc. to identification label.
AC ±10 %
Power fuse:
T1A L250 V
Nominal power:
350 VA
Maximum start-up current (5 s)
25 A
Operating temperature range:
+15 °C to +28 °C
Operating temperatures:
-5 °C up to a room temperature of approx. +28 °C
Relative humidity:
20 to 80 % - non-condensing
Permissible temperature range
during storage and transport:
-29 °C to +50 °C
Permissible humidity range
during storage and transport:
10 to 85 % - non-condensing
IEC 1010 classification:
Protection class 1
Pollution degree
2
Operating elevation:
up to max. 2000 m NN
IP protection class (IEC 60529):
IP20
Refrigeration unit
Refrigeration capacity*:
110 W
Safety factor:
3
Refrigerant:
R 134a
Compressor oil:
180 cm
3
Emkarate RL 15 s, ICI
*according to CECOMAF, fluid temperature: 55 °C, evaporation temperature: -25 °C
Dimensions and weight
Width:
360 mm
Depth:
650 mm
Height:
360 mm
Weight:
23 kg
Instrument type Power supply voltage
Coolant
EG1150 C-1
230 V / 50 Hz
80 g ± 5 g
EG1150 C-3
120 V / 60 Hz
80 g ± 5 g
EG1150 C-8
240 V / 50 Hz
80 g ± 5 g
EG1150 C-11
100 V / 50 - 60 Hz
65 g ± 5 g
3. Instrument Components and Specifications