4. Instrument Setup
4.2 Delivery range
The basic equipment for the Leica EG1150 C cold plate contains the
following components:
Order No.
• 1 basic unit Leica EG1150 C
• 1 power cable
see
p. 15
only for Australia:
• 1 power cable EU
14 0411 13558
• 1 series transformer, 230/240 V
14 0469 31998
• 1 replacement fuse T 1.0 A, 6.3x32 mm
14 6943 01001
• 1 Instructions for Use German/English (+ CD)
14 0388 80002
• Optional: Unit cover (acrylic)
4.1 Location conditions
• Stable, vibration-free laboratory table with horizontal, flat table top, as
far as possible vibration-free ground.
• No direct sunlight or strong temperature fluctuations. Room tempera-
ture consistently between +15 °C and +28 °C.
• Relative air humidity maximum 80 %, non-condensing.
• The instrument should be set up in such a way that the air circulation
is not impaired.
At a room temperature of > +28 °C, the working temperature of the
cold plate of -5 °C may not be reached at all points.
To ensure proper function, there must be gap of at least 15 cm
behind the instrument. Failure to observe this distance may result
in serious damage to the refrigeration unit of the device. The instru-
ment should not be operated in hazardous locations.
Please compare the delivered components against the packing list
and your order.
Should there be any discrepancy, please contact the Leica distributor
handling your order.