4 netzanschluss, Power supply, Instrument setup – Leica Biosystems EG1150 C User Manual
Page 15

15
Leica EG1150 C
4. Instrument Setup
4.4 Power supply
The Leica EG1150 C refrigeration unit requires a specific voltage and
frequency (refer to
"Technical data" chapter
), and is therefore always
delivered with a power cable that fits the instrument.
Please observe the following notes to prevent damage to the unit:
• Connect the power cord plug (7) to the connecting port (6).
• Plug the power cord into the wall outlet.
Fig. 6
6
7
Only one of the following network cables may
be used to connect the instrument:
Application area
Order No.
Australia
14 0411 32565
Europe:
14 0411 13558
Italy:
14 0411 13561
Switzerland:
14 0411 13560
United Kingdom:
14 0411 27822
USA, Canada, Japan:
14 0411 13559
Brazil
14 0411 47869
Before connecting the instrument to the power supply, it is manda-
tory to check whether the voltage specified on the identification la-
bel (rear side) matches the actual voltage values at the installation
location.
If this is not the case, the connection must not be made!
The unit must be connected only with the supplied power cable and
only to a grounded power receptacle.
Do not use an extension cord!