Leica VT1000 S - Vibrating-blade microtome

5. Operation

5.3 The operating elements and their function - VT1000 S

Scale setting mm/s

0 0.00

0.5 0.025

1 0.050

2 0.075

3 0.125

4 0.175

5 0.225

6 0.40

7 0.65

8 0.90

9 1.30

10 2.50

10-speed rotating potentiometer with scale
Function:

Continuous knife feed adjustment from 0.05 -

2.5 mm/s:

Knife return stroke is performed at constant speed

of 5 mm/s.

The additional locking lever (lever in 12 o'clock

position) prevents the speed setting from be-

ing accidentally changed while sectioning is in

progress.

SPEED

Rotary knob with scale from 0 to 10

Function:

Continuous adjustment of knife sectioning fre-

quency (vibration) from 8 - 100 Hz.

Scale setting Hz

0

0

0.5

8

1 10

2 20

3 30

4 40

5 50

6 60

7 70

8 80

9 90

10 100

FREQ

FREQ

Locking lever

SPEED

Caution: Practice working with the controls without a blade holder inserted. Only insert

the knife holder when you are completely familiar with all control functions.

Fig. 14

Fig. 15