6.
Malfunctions: Meaning and Troubleshooting
Error messages/symptoms
Sources of error
Troubleshooting
- Audible warning signal.
- Error code E0.1xx is displayed.
xx - there are several error codes,
00 - there is only one error code.
- Button(s) jammed or defective.
- Locking function /REV or REV/
FORW button defective.
- Error on the UP/DOWN button;
DOWN locking function.
- Push the button several times to
unlock;
have defective button replaced
by the Technical Service.
Error code E0.200 is displayed.
- Feed mechanism defective.
- Switch off the instrument; call
the Technical Service.
- Error code E0.300 is displayed.
- Important electronic component
defective.
- Switch off the instrument; call
the Technical Service.
- Error code E0.400 is displayed.
- Feed motor defective.
- Switch off the instrument; call
the Technical Service.
Error code E.05xx is displayed.
- Light barrier error (forward
feed)
- Switch off the instrument; call
the Technical Service.
- Audible warning signal.
- Error code E0.600 is displayed.
- Light barrier error (section
thickness feed)
- Switch off the instrument; call
the Technical Service.
- Audible warning signal.
- Error code E0.700 is displayed.
for approx. 2 secs.
- Software detected severe hard-
ware fault.
- Switch off the instrument; call
the Technical Service.