6.

Malfunctions: Meaning and Troubleshooting

Error messages/symptoms

Sources of error

Troubleshooting

- Audible warning signal.

- Error code E0.1xx is displayed.

xx - there are several error codes,

00 - there is only one error code.

- Button(s) jammed or defective.

- Locking function /REV or REV/

FORW button defective.

- Error on the UP/DOWN button;

DOWN locking function.

- Push the button several times to

unlock;

have defective button replaced

by the Technical Service.

Error code E0.200 is displayed.

- Feed mechanism defective.

- Switch off the instrument; call

the Technical Service.

- Error code E0.300 is displayed.

- Important electronic component

defective.

- Switch off the instrument; call

the Technical Service.

- Error code E0.400 is displayed.

- Feed motor defective.

- Switch off the instrument; call

the Technical Service.

Error code E.05xx is displayed.

- Light barrier error (forward

feed)

- Switch off the instrument; call

the Technical Service.

- Audible warning signal.

- Error code E0.600 is displayed.

- Light barrier error (section

thickness feed)

- Switch off the instrument; call

the Technical Service.

- Audible warning signal.

- Error code E0.700 is displayed.

for approx. 2 secs.

- Software detected severe hard-

ware fault.

- Switch off the instrument; call

the Technical Service.