Leica VT1000 S - Vibrating-blade microtome

5. Operation

• The instrument is designed for indoor use only.
• The power plug must be freely and easily accessible.
• Power supply at a distance no greater than the length of the power cable

(3m) – an extension cable must not be used.

• Level installation location,
• Substrate as free of vibration as possible,
• Relative humidity should not exceed 80 %
• Room temperature consistently between +5 °C and +40 °C
• Avoid vibrations, direct sunlight, and large temperature fluctuations!

The instrument MUST be connected to a grounded power socket. Use only a provided

power cable that is intended for the local power supply.

5.1 Installation site requirements

The place of installation must meet the following requirements:

The instrument may not be operated in hazardous locations.