7. Troubleshooting
7.1 Malfunctions
+
Confirm elimination
of error
+
If the printer malfunctions during printing, a corresponding error
code is displayed and simultaneously the LED in the ERROR button
starts flashing.
How to eliminate the problem:
• Determine the cause of the error—see error list (
Chapter 7.3
).
• Remove obstruction(s); open lid if required.
• Important!
Remove all slides which are still in the chute, in or next to the specimen
slide carrier or in the drying module!
These specimen slides should not be used again.
• Reclose the hood and press the ERROR button to confirm to the printer
that the source of the error has been eliminated.
• The printer then verifies whether all processing paths are unobstructed
and whether the source of the error has been eliminated.
• If there are still some obstructions left, or if the source of error has not
been thoroughly eliminated, the printer displays another error message.
• Interrupted print jobs are resumed where previously left off.
• If an error message is displayed several times although all possible
causes have been eliminated, a Reset should be carried out.
Reset:
• Press and release LOADED and ERROR simultaneously.
• A Reset restores the printer to the same state as directly after switching
on. All print jobs in the print queue are deleted.
• If the same error continues to be displayed even after a Reset, switch the
printer off via the power switch (back panel) and, after a short waiting
period of approx. 30 sec., switch it back on. If this does not eliminate the
problem either, call Leica Technical Service.