5. Operation
5.3 Alarm functions
The Leica IP S is equipped with two different alarm functions:
Instrument alarm
The printer has a beeper that emits acoustic signals indicating important
instrument states and functions.
• Upon pressing the button:
1 short beeping sound
• Magazine empty/tray stack full:
2 short beeping sounds
• In case of error:
5 short beeping sounds
• When ending the head cleaning:
5 short beeping sounds
The beeper can be deactivated
by means of the DIL switches
at the back panel of the printer.
To deactivate the beeper, push
the switch at the very bottom
(60) to the right
(in the direction of the red
arrow in
Fig. 45)
.
Fig. 45
60
Remote alarm
Additionally, an alarm can be installed outside the room in which the printer
works.
• The remote alarm device is connected to the printer via a 3.5 mm-
diameter jack connector (optional accessory) that is inserted into socket
(56).
• The remote alarm is triggered if no power is supplied to the printer or if
the power switch at the back panel of the printer is switched off.
56
Back panel of instrument
The remote alarm device connected to the instrument must be rat-
ed at less than 100 mA. A maximum voltage of 24 V DC must not be
exceeded.
For details on how to connect a remote alarm device to the Leica
Printer IP S, please contact your local Leica sales office or the
manufacturer directly.