Kds ezflow 2040, Operation settings and procedures – KD Scientific EZFLOW 2040 User Manual
Page 25
![background image](/manuals/564072/25/background.png)
23
KDS EZFlow 2040
Operation Settings and Procedures
(Continued)
10. FACTORY RESET
MENU
ENTER
SET
ENTER
SET
Screen 23
1)
Press
to enter MODE SETTING
status.
2)
Press
or
to select item No.9
FACTORY RESET.
3)
Press
to enter FACTORY RESET
menu. (Screen 23 ).
4)
Select CONFIRM, press
to reset
the pump.
WARNING: Once pump is reset, all settings will return to factory default as following:
1)
All preset data and pre-programmed infusion modes will be cleared.
2)
The occlusion level will be reset to medium.
3)
The accuracy calibration will be reset to 0%
4)