Kds ezflow 2040, Quick operating instructions – KD Scientific EZFLOW 2040 User Manual
Page 15
13
KDS EZFlow 2040
Quick Operating Instructions
1)
Press ON/OFF switch on the back of the pump to turn on the device.
2)
Close roller clamp on set, spike and hang fluid container. Fill drip chamber to
fill line.
3)
Open roller clamp to fill all the set with fluid. Close roller clamp.
4)
Open door, install set. Close door. Open roller clamp.
5)
Press
once to enter the infusion rate setting screen, press or hold
or
to set desired infusion rate. Press
once to confirm the desired rate and
jump to volume setting.
6)
Press or hold
or
to set desired volume to be infused. Press
to con-
firm the setting and return to the main screen.
7)
Attach the infusion set to vein-puncture device.
8)
Hold
to prime the infusion set.
9)
Connect the vein-puncture device to the patient. Press
to begin infusion.
10) To stop infusion, press
11) To restart or clear the total infused volume, press ON/OFF switch on the back of
the pump to restart the pump (minimum 3 seconds interval).
ALARMS
•
To silent alarm, press
•
Once Flash light and audio alarm is triggered, please note the alarm messages
and correct cause of alarm on the screen.