Kds ezflow 2040, Operation settings and procedures – KD Scientific EZFLOW 2040 User Manual
Page 21
![background image](/manuals/564072/21/background.png)
19
KDS EZFlow 2040
Screen 16
IV SET
Drop Rate Range
Equivalent Flow Rate
Operation Settings and Procedures
(Continued)
7. ML OR DROP SETTING
(Continued)
Screen 17
Screen 18
5)
If drop/min is selected, press
to
enter the SELECT IV SET menu as shown
on Screen 17 and Screen18 for different
IV set setting.
6)
Please verify the IV set type on the IV
set’s packaging then press
or
to find the correct IV set setting.
6)
Press
to confirm selection and
return to MODE SETTING screen.
7)
Press
to return to main screen
ENTER
SET
ENTER
SET
EXIT
CLEAR
WARNING: Failure to select correct IV set will result inaccurate infused volume counting
under drop/min mode.
IMPORTANT: Due to the design of the IV set, the minimum flow rate and maximum drop
rate are varied for different IV set under drop/min mode, check the following conversion
chart:
10 Drops/ml:
1/200 D/min
6/1200ml/h
15 Drops/ml:
1/200 D/min
4/800ml/h
20 Drops/ml:
1/200 D/min
3/600ml/h
60 Drops/ml:
1/180 D/min
1/180ml/h
IMPORTANT: For 60 Drops / ml IV set, the maximum drop rate is 1580 D/min; the maxi-
mum flow rate is 1580ml/h.