Kds ezflow 2040, Operation settings and procedures – KD Scientific EZFLOW 2040 User Manual
Page 24
22
KDS EZFlow 2040
Operation Settings and Procedures
(Continued)
9. OCCLU. LEVEL (OCCLUSION LEVEL)
MENU
ENTER
SET
Screen 21
Screen 22
HIGH: If the pump is not sensitive enough to trigger occlusion alarm with certain I.V sets,
HIGH occlusion sensitivity level should be selected.
MEDIUM: MEDIUM occlusion sensitivity level is the factory default setting. This occlusion
sensitivity level is preferred for most scenarios.
LOW: If the pump is too sensitive enough to trigger occlusion alarm, LOW occlusion sensitivi-
ty level shall be selected.
LOWEST: If the pump is still too sensitive and frequently sets off occlusion alarm with other
I.V sets, LOWEST occlusion sensitivity level shall be selected.
WARNING: The occlusion alarm will not be triggered under LOWEST OCCLUSION SEN-
SITIVITY LEVEL for most pump use IV sets.
1)
Press
to enter MODE SETTING
status.
2)
Press
to item No.8 OCCLU. LEVEL
(OCCLUSION LEVEL).
3)
Press
to enter ADJ. OCCLUSION
SENSITIVITY menu (screen 21 and
screen 22). There are 4 levels of
occlusion sensitivity to adjust: