Kds ezflow 2040, Operation settings and procedures – KD Scientific EZFLOW 2040 User Manual
Page 23
![background image](/manuals/564072/23/background.png)
21
KDS EZFlow 2040
ENTER
SET
ENTER
SET
Screen 20
Operation Settings and Procedures
(Continued)
8. ADJ. I.V SET (ACCURACY CALIBRATION
Continued
)
MANUAL ADJUSTMENT (For ml/h or drop/min mode)
1)
Press
to choose manual
calibration. (screen 20)
2)
Press
or
to change calibration
volume in percentage.
3)
Press
to confirm the adjustment
and return to the MODE SETTING status.
IMPORTANT : The manual adjustment is in effect at the current flow rate from 0.1 to 1000
ml/hr ONLY. The pump will NOT respond to the manual adjustment if the current flow rate is
set from 1001 to 1200 ml/hr.