Kds ezflow 2040, Start up – KD Scientific EZFLOW 2040 User Manual
Page 13
11
KDS EZFlow 2040
Start Up
IMPORTANT: For the first time use, user MUST read this manual carefully.
Screen 1
1)
Connect the power cord firmly to the
pump and electrical outlet to charge
internal batteries. Pump can still be used
while batteries are charging. During the
operations, if the power cord is discon-
nected, pump will automatically switch to
internal battery power and will keep a
continuous normal operation. Batteries
charged over 8 hours can support con-
tinuous operation for over 3 hours at
25ml/h infusion rate.
2)
Pump has self-diagnostic function. The
pump will initiate a self-diagnostic pro
gram to go through key components
every time the pump is turned on. After
self-diagnosis is completed successfully,
user will see a welcome screen with
company logo, followed by the main
screen (screen 1). Then this pump is
ready for use.