6 Technical data

Galvanic isolation

Without galvanic isolation between sensor
and output

Leakage resistance

> 100 M

Ω

at DC 100 V measured at room

temperature between connection
terminals and case

Reverse polarity
protection

Yes

Voltage supply error

±0.01 % per V deviation from 24 V

The % value refers to the range end value
20 mA

Environmental influences

Ambient temperature
range for head

-30 to +85 °C

Storage temperature
range

-30 to +90 °C

Ambient
temperature error

±(15 ppm/°C × (range end value +200) +

50 ppm/°C × measuring range set) ×

Δυ

Δυ

= deviation of ambient temperature

from the reference temperature

Calibration/reference
conditions

DC 24 V at 25 °C ±5 °C (77 °F ±9 °F)

Climatic conditions

According to IEC 60068-2-30
(rel. humidity

95 % with condensation)

Vibration strength

According to IEC 60068-2-6
(according to GL characteristics)

Electromagnetic
compatibility (EMC)
- interference emission
- interference immunity

DIN EN 61326
class B
industrial requirements

Protection type

IP67 according to DIN EN 60529
with machine connector plugged in