JULABO F38-ME Beer Forcing Test Refrigerated/Heating Circulating Bath User Manual
Page 65
ME
65
in commands:
Asking for parameters or temperature values to be displayed.
Command
Parameter
Response of circulator
version
none
Number of software version (V X.xx)
status none
Status message, error message (see page 66)
in_pv_00
none
Actual bath temperature.
in_pv_01
none
Heating power being used (%).
in_pv_02
none
Temperature value registered by the external Pt100 sensor.
in_pv_03
none
Temperature value registered by the safety sensor.
in_pv_04 none Setpoint temperature of the excess temperature protection
in_sp_00
none
Working temperature „SETPNT 1“
in_sp_01
none
Working temperature „SETPNT 2“
in_sp_02
none
Working temperature „SETPNT 3“
in_sp_03 none High temperature warning limit
„OVERTEMP“
in_sp_04 none Low temperature warning limit
„SUBTEMP“
in_sp_07
none
Pump pressure stage
in_par_01
none
Te - Time constant of the external bath.
in_par_02
none
Si - Internal slope
in_par_03
none
Ti - Time constant of the internal bath.
in_par_04
none
CoSpeed - Band limit (max. difference between the
temperatures in the internal bath and external system).
in_par_05
none
Factor pk/ph0:
Ratio of max. cooling capacity versus max. heating capacity
in_par_06
none
Xp control parameter of the internal controller.
in_par_07
none
Tn control parameter of the internal controller.
in_par_08
none
Tv control parameter of the internal controller.
in_par_09
none
Xp control parameter of the cascade controller.
in_par_10
none
Proportional portion of the cascade controller.
in_par_11
none
Tn control parameter of the cascade controller.
in_par_12
none
Tv control parameter of the cascade controller.
in_par_13
none
Adjusted maximum internal temperature of the cascade
controller.
in_par_14
none
Adjusted minimum internal temperature of the cascade
controller.
in_par_15
none
Band limit (upper)
in_par_16
none
Band limit (lower)