beautypg.com

JULABO F38-ME Beer Forcing Test Refrigerated/Heating Circulating Bath User Manual

Page 52

background image

Menu functions

52

9.7.3.

CALIBRATION TYPE: 1 -/ 2 -/ 3 POINT

A >1-point<, >2-point< or >3-point< calibration can be carried out.

First geometrically define the location for calibration (measuring point CT),
then determine the temperature values of the calibration points.
The type of calibrations also determines the number of the following pairs of
values indicated on the LCD DIALOG-DISPLAY.

Pairs of values:

TMPVAL X:

Circulator temperature 1 or 2 or 3 (actual value TT )

The actual temperature of the bath is simultaneously saved with the
“calibration value“ >CALVAL< and can be indicated for control purposes
(value does not flash).


CALVAL X:

Calibration temperature 1 or 2 or 3 (actual value CT )

The „calibration value“ is determined with a temperature measuring
device and saved under menu item >CALVAL<.

(value flashes

)