8.3.

Over and Sub temperature warning function

Over temperature

Sub temperature

Recommendation:

Settings of the forcing
test

OVERTEMP 65,0

°C

SUBTEMP -5,0

°C

If the observance of a working temperature value >SETP< has to be
supervised for a sensitive temperature application, then set over and sub
temperature warning values. In the example below the SETPOINT 85 °C is
surrounded by the values OVERTMP 87 °C and SUBTEMP 83 °C. The
electronics immediately register if the actual temperature breaches one of
the set limit values. The resulting reaction is defined in a further menu item.
(See chapter 8.2.

)

C

1. Press the key

.

2. By pressing the

or

key select the menu >OVERTMP< or

>SUBTEMP<.

3. Press

the

key. The integer digits flash

4. Change the values to 87. °C and/or 83. °C by pressing the

and

key and confirm with the

key.

The decimal digits flash and can be adjusted if desired.

Confirm once more by pressing the

key.

See above examples.

The warning functions are only activated if the actual bath temperature
remains within the set limit values for 3 seconds after switch-on.

Recommendation:

Set the over temperature warning value >OVERTMP< 5 °C to 10 °C above
the working temperature setpoint.

Set the sub temperature warning value >SUBTMP< 5 °C to 10 °C below the
working temperature setpoint.