JULABO F38-ME Beer Forcing Test Refrigerated/Heating Circulating Bath User Manual
Page 27
ME
27
8.3.
Over and Sub temperature warning function
Over temperature
Sub temperature
Recommendation:
Settings of the forcing
test
OVERTEMP 65,0
°C
SUBTEMP -5,0
°C
If the observance of a working temperature value >SETP< has to be
supervised for a sensitive temperature application, then set over and sub
temperature warning values. In the example below the SETPOINT 85 °C is
surrounded by the values OVERTMP 87 °C and SUBTEMP 83 °C. The
electronics immediately register if the actual temperature breaches one of
the set limit values. The resulting reaction is defined in a further menu item.
(See chapter 8.2.
)
C
1. Press the key
.
2. By pressing the
or
key select the menu >OVERTMP< or
>SUBTEMP<.
3. Press
the
key. The integer digits flash
4. Change the values to 87. °C and/or 83. °C by pressing the
and
key and confirm with the
key.
The decimal digits flash and can be adjusted if desired.
Confirm once more by pressing the
key.
See above examples.
The warning functions are only activated if the actual bath temperature
remains within the set limit values for 3 seconds after switch-on.
Recommendation:
Set the over temperature warning value >OVERTMP< 5 °C to 10 °C above
the working temperature setpoint.
Set the sub temperature warning value >SUBTMP< 5 °C to 10 °C below the
working temperature setpoint.