beautypg.com

Remote control – JULABO F38-ME Beer Forcing Test Refrigerated/Heating Circulating Bath User Manual

Page 62

background image

Remote control

62

/

Control output

The

connector may be used for control of JULABO refrigerated

circulators or as output for alarm messages.

1

3

2

4

5

Circuit:

Operation

= relay powered

Alarm

= relay not powered

Pin assignment:
Pin

Signal

1

+24 V (I max. current 25 mA)

2

0 V

3

Alarm relay

4

Reserved - do not use!

5 Cooling

pulse
























12. Remote

control

12.1.

Setup for remote control

9

6

5

1

RS232C

1. Check the interface parameters for both interfaces (on circulator and PC)

and make sure they match.
(Serial interface refer to page 48)

2. In the menu > MENU CONFIG < set the menu item > REMOTE < to

> ON < .

3. Connect both units with an interface cable..

Like all parameters which can be entered through the keypad, interface
parameters are stored in memory even after the circulator is turned off.