37

9. Set MENU Program to >Mode Forcier<

See page 32


10. Start the forcing test

As factory-setting the „Forcing test 60 °C“ is stored

Measuring cycle:RUNS

1 flow

(48 h)

Warm phase

W-TEMP

60 °C

Duration

W-TIME

24 h

Cool phase

C-TEMP

0 °C

Duration

C-TIME

24 h


Setting of other values see page 32 FORCING-Test Program

Start-Menu

Press the

key if a parameter is to be retained.

Correction function for parameters or values (prior to OK)

Menu level 1

> RUNS < Number of repetitions of the warm/cool phase 1 ... 10

> GO <

Start the forcing test (NOW)

Level 2

Parameter level

Level 2


11. Program sequence