Start-menu – JULABO F38-ME Beer Forcing Test Refrigerated/Heating Circulating Bath User Manual
Page 29

ME
29
9.1.
MENU PROGRAM – START
9.1.1.
START Mode Forcing test
Start-Menu
This menu will start a previously set program.
Requirements:
1. Create a program. (refer to chapter >MENU PROGRAM<)
2. Return to the Start-MENU and confirm the desired setting of each
MENU item with the key
Press the
key if a parameter is to be retained.
Correction function for parameters or values (prior to OK)
Menu level 1
> RUNS < Number of repetitions of the warm/cool phase 1 ... 10
> GO <
Start the forcing test (NOW)
Level 2
Parameter level
Set number of runs with
and
example: 1 run
Confirm >NOW< with the
key and the program
will start immediately
9.1.2.
START Mode Standard
Start-Menu
This menu will start a previously set program.
Requirements:
3. Create a program. (refer to next chapter)
4. Return to the Start-MENU and confirm the desired setting of each
MENU item with the key
5. Set a start time (>TIME< >DATE< >YEAR<) if the program is to be
started by the internal timer.
Menu level 1
> STEP<
Program start at section 1 … 10
> RUNS <
Number of repetitions 1 ... 99
> END<
Status at end of program (STDBY/SETPNT)
Standby or last setpoint
> GO <
Time of start (NOW/TIMER)