beautypg.com

Start-menu – JULABO F38-ME Beer Forcing Test Refrigerated/Heating Circulating Bath User Manual

Page 29

background image

ME

29

9.1.

MENU PROGRAM – START

9.1.1.

START Mode Forcing test

Start-Menu

This menu will start a previously set program.

Requirements:

1. Create a program. (refer to chapter >MENU PROGRAM<)

2. Return to the Start-MENU and confirm the desired setting of each

MENU item with the key

Press the

key if a parameter is to be retained.

Correction function for parameters or values (prior to OK)

Menu level 1

> RUNS < Number of repetitions of the warm/cool phase 1 ... 10

> GO <

Start the forcing test (NOW)

Level 2

Parameter level

Set number of runs with

and

example: 1 run

Confirm >NOW< with the

key and the program

will start immediately


9.1.2.

START Mode Standard

Start-Menu

This menu will start a previously set program.

Requirements:

3. Create a program. (refer to next chapter)

4. Return to the Start-MENU and confirm the desired setting of each

MENU item with the key

5. Set a start time (>TIME< >DATE< >YEAR<) if the program is to be

started by the internal timer.

Menu level 1

> STEP<

Program start at section 1 … 10

> RUNS <

Number of repetitions 1 ... 99

> END<

Status at end of program (STDBY/SETPNT)

Standby or last setpoint

> GO <

Time of start (NOW/TIMER)