JULABO F38-ME Beer Forcing Test Refrigerated/Heating Circulating Bath User Manual
Page 36

Menu functions
36
9.2.3.
The easy way to the forcing test
Notice:
On the forcing test avoid low level alarm CODE 1.
The program is cancelled. On a restart the count of the cycles (CYCLE)
restarts from zero.
Refill the bath liquid in time!
1. Installation
see page 17.
2. Power connection
see picture „Rear view“ page 20
3. Filling / Bath fluid
Filling: see page 19
4. Insert the the product
Put the stainless steel insert cage with beer bottles into the bath tank
and refill the fluid. Maximum filling height approx. 25 mm under the bath
edge.
5. Put on the bath cover
In order to reduce the loss of fluid due to evaporation of water, in case
of long-term operation.
The circulator ME is provided with an early-warning system for
low level, which activates at a filling height of 65-70 mm below the
bath edge. For measuring cycles, which last for several days the
liquid level should be peered once a day, in order to not to
endanger the course of the forcing test.
6. Switching on:
Circulator and cooling machine may be turned on and off with separate
mains switches (1a, 1b).
7. Safety installations, warning functions
Set the following value in the MENU
Excess temperature protection see page 25
Set >SAFETMP< to 70 °C
Over and Sub temperature warning functions see page 27
Set >OVERTMP< to 65 °C
Set >SUBTMP< to -5 °C
8. Setting the test temperature
Select e.g. SETPNT 1 in the MENU
and set 20 °C.
see page 22