JULABO F38-ME Beer Forcing Test Refrigerated/Heating Circulating Bath User Manual

Page 38

background image

Menu functions

38

Program sequence:

Example forcing test 60 °C

As factory-setting the „Forcing Test 60 °C“ is stored

Measuring cycle: RUNS

1 repetition

Warm phase

W-TEMP

60 °C

Duration

W-TIME

24 h

Cool phase

C-TEMP

0 °C

Duration C-TIME

24

h

In the course of the program the VFD-DISPLAY shows messages
concerning the process of the forcing test in line 2 and 3.

In line 3 two messages are shown alternately.
Message 1 always indicates the program status (phase of the forcing test),
message 2 indicates the status >RUN< (Test runs).

Message 2 can be changed by the editing keys

.The remaining

time of a phase or >E ------< can still be chosen.
(Message >E ------<, temperature value of the external temperature sensor
is not important for the forcing test.)

Action

diagram:

1.

Set measuring temperature :

Select in the MENU

e.g. SETPNT 1 and set 20 °C.

(see page 22)

2. Start forcing test (see page 29)