Operating procedures

20

6. Operating

procedures

6.1. Power

connection

Caution:

 Only connect the unit to a power socket with earthing contact (PE – protective

earth)!

 The power supply plug serves as safe disconnecting device from the line and must

be always easily accessible.

 Never operate equipment with damaged mains power cables.

 Regularly check the mains power cables for material defects (e.g. for cracks).

 We disclaim all liability for damage caused by incorrect line voltages!

Check to make sure that the line voltage matches the supply voltage
specified on the identification plate. Deviations of ±10 % are permissible.

 Connect the circulator with mains power cable (20a) to the mains

outlet (21).

 Connect the control cable (15a) between the connectors  (15, 15b).
 Connect the refrigerated circulator with mains power cable (20b) to

the mains socket.

6.2. Switching

on / Start - Stop

6.2.1. Switching

on the circulator

Switching on:

 Turn on the mains power switch (1).

The unit performs a self-test.
Then the software version (example: V 1.xx) appears. The display
OFF“ or „R OFF“ indicates the unit is ready to operate.

The circulator enters the operating mode activated before switching
the circulator off:
keypad control mode (manual operation)
or
remote control mode (operation via personal computer).