JULABO F38-ME Beer Forcing Test Refrigerated/Heating Circulating Bath User Manual

Page 54

background image

Menu functions

54

An ATC calibration is to be carried out.
Set to >NO<

 The parameter flashes, switch by pressing

and

.

 The parameter flashes, switch by pressing

and

.

A >3-point< calibration is carried out.

The value >TMPVAL< is only indicated

In addition the measured value >CALVAL X< is saved
during the following step

 Integer digits flash, set by pressing

(79) +

 Decimal digits flash, set by pressing

(70) +

The first of 3 points is calibrated.

Return to 2. Set working temperature value SETPNT: 120.00 °C

The value is only indicated

 Integer digits flash, set by pressing

(119) +

 Decimal digits flash, set by pressing

(50) +

The second of 3 points is calibrated.

Return to 2. set working temperature value SETPNT: 160.00 °C

The value is only indicated.

 Integer digits flash, set by pressing

(159) +

 Decimal digits flash, set by pressing

(30) +

The 3-point calibration is completed

5. Return to standard operation

 Set >YES< after calibration.

(Standard operation)