INFICON SKY CDGxxxD (DeviceNet) User Manual
Page 25
tira51e1 (2009-01) CDGxxxD.cp
25
Attribute
ID
Access rule NV/
V
Name
DeviceNet
data type
Description of attribute
Semantics of values
3
→
"Semantics"
NV Data type USINT
Determines the data type
of value and all related
attributes as specified in
this table.
→ "Semantics" section below
[default] = INT
4
→
"Semantics"
NV Data units USNIT
Determines the units con-
text of value and all related
attributes.
[default] = Counts
5 Get
V
Reading
valid
BOOL
Indicates that the value
attribute contains a valid
value.
0 = invalid
1 = valid (always set)
6
Get
V Value
INT or REAL
Analog input value.
The corrected, converted, calibrated
final value of the sensor.
→ "Semantics" section
7
Get
V Status
BYTE
Alarm and warning state of
this object instance.
= 0
10
Get
NV Full scale
INT or REAL
The value of full scale for
the sensor.
[default] = maximum allowable value
for the data type
25
Set
NV Safe state USINT
Specifies the behavior for
the value for states other
than execute.
→ "Semantics" section below
[default] = 0
26
Set
NV Safe value INT or speci-
fied by data
type
The Value to be used for
safe state = safe value
→ "Semantics" section below
[default] = 0
32
Get
NV Overrange INT or REAL
Specifies the highest valid
value.
The value above which attribute
reading valid is set to invalid.
[default] = maximum allowable value
for the data type
33 Get
NV
Under-
range
INT or REAL
Specifies the lowest valid
value.
The value below which attribute
reading valid is set to invalid.
[default] = minimum allowable value
for the data type
94 Get
V
Sensor
warning
Structure of
byte
Bit definitions of sensor
warnings.
default = 0
→ "Semantics" section below
95 Get
V
Sensor
alarm
Structure of
byte
Bit definitions of sensor
alarms.
default = 0
→ "Semantics" section below
96 Get
V
Status
extension
Byte
Bit-mapped byte providing
additional status bits of an
S-Analog sensor instance.
default = 0
→ "Semantics" section below
99 Get
V
Subclass
UINT Identifies a subset of addi-
tional instance attributes,
services and behaviors.
3 = Capacitance Diaphragm Gauge
→ Instance 1
→ Instance 1
→ Instance 1
→ Instance 1
→ Instance 1
4.4.3.6 Instance Attributes of
Instance 2 / Atmosphere
Pressure Sensor
4.4.3.7 Semantics of S-Analog
Sensor Instance 2
Data type
Data units
Value
Safe state
Safe value