INFICON SKY CDGxxxD (DeviceNet) User Manual
Page 20
20
tira51e1 (2009-01) CDGxxxD.cp
Following is the instance 1 with the subclass extension of the capacitance mano-
meter (diaphragm gauge) part of the gauge. Many different types of gauges are
available. The behavior of all gauges is the same, but the fullscale value is
different.
Attribute
ID
Access rule
NV/
V
Name
DeviceNet data
type
Description of
attribute
Semantics of values
3 Set
/Cond.
→ below
NV Data type
USINT
Determines the
data type of value
and all related
attributes as speci-
fied in this table.
→ "Semantics" section below.
Int
≙
C3
h
[default]
float
≙
CA
h
4 Set
→ below
NV Data units
UINT
Determines the
units context of
value and all re-
lated attributes.
→ "Semantics" section below.
Supported Values:
Counts
≙
1001
h
[default]
mbar
≙
1308
h
Torr
≙
1301
h
Pascal
≙
1309
h
5
Get
V Reading valid
BOOL
Indicates that the
value attribute con-
tains a valid value.
0 = invalid
1 = valid
(invalid: e.g., not warmed up
yet)
6
Get
V Value
INT or specified
by data type
Analog input value The corrected, converted, cali-
brated final value of the sensor.
→ "Semantics" section below.
7
Get
V Status
BYTE
Alarm and warning
state of this object
instance.
→ "Semantics" section below.
8
Set
NV Alarm enable
BOOL
Enables the setting
of the alarm status
bits
0 = disable [default]
1 = enable
9
Set
NV Warning enable
BOOL
Enables the setting
of the warning
status bits
0 = disable [default]
1 = enable
10
Get
NV Full Scale
INT or specified
by data type
The value of full
scale for the sen-
sor.
The value of attribute value
corresponding to the full scale
calibrated measurement of the
sensor.
[default] = maximum allowable
value for the data type
11
Get
NV Offset-A data type
USINT
Determines the
data type of
attribute offset-A
[default] = INT (C3
h
)
12 Set
NV
Offset-A Specified
by
attribute 11
An amount added
prior to gain to
derive value
This value could be read at the
measurement device itself after
execution of zero adjust service.
0 = [default]
17
Set
NV Alarm trip point
high
Determines
the
value above which
an alarm condition
will occur
[default] = Maximum value for
its data type.
18
Set
NV Alarm trip point low INT or specified
by data type
Determines the
value below which
an alarm condition
will occur
[default] = Minimum value for its
data type.
19
Set
NV Alarm hysteresis
INT or specified
by data type
Determines the
amount by which
the value must
recover to clear an
alarm condition
[default] = 0
4.4.3.1 Instance Attributes of
Instance 1