INFICON BCG450-SP ATM to Ultra-High Vacuum Triple Gauge (Profibus) User Manual
Page 40

40
tira41e1 (2005-06) BCG450SPv1.cp
ID
Name
Structure Data type
Bytes
Access
Store
101
Block Type
Simple
Octet string
4
2_R
102
Status Extension
Simple
UINT8
1
1/2_R
V
103
Sensor Alarm
Struct
Array of 2 bytes
2
1/2_R
V
104
Sensor Warning
Struct
Array of 2 bytes
2
1/2_R
V
According to the table in Appendix A, the Block Type ID has the value "14".
This parameter indicates whether the overrange or underrange of the diaphragm
gauge device instance is exceeded.
Bit7
Bit6
Bit5
Bit4
Bit3
Bit2
Bit1
Bit0
0
0
0
0
0
Underrange
Exceeded
Overrange
Exceeded
Reading
Invalid
If the instance is operated in its overrange or underrange, the corresponding bit
and additionally the bit "Reading Invalid" is set.
If an error occurs, the bit "Reading Invalid" as well as the corresponding error bits
in Device Block (ID 26, 27, 28 Device Block) are set.
This parameter indicates the detectable errors occurring in connection with the
diaphragm gauge measuring part. The present implementation allows detection of
one error.
Bit7
Bit6
Bit5
Bit4
Bit3
Bit2
Bit1
Bit0
Byte 0
0
0
0
0
0
0
0
0
Byte 1
0
0
0
0
0
0
Electronics Failure
1)
0
1)
Electronics Failure includes a sensor error.
The Sensor Alarm bits defined here are copied into the Device Block ID 27 in the
"Device Exception Detail Alarm" range of the the capacitance diaphragm gauge
measuring part.
This parameter indicates the detectable warnings occurring in connection with the
diaphragm gauge measuring part. The present implementation allows detection of
one warning.
Bit7
Bit6
Bit5
Bit4
Bit3
Bit2
Bit1
Bit0
Byte 0
0
0
0
0
0
0
0
0
Byte 1
0
0
0
0
0
0
Electronics Warning
0
The Sensor Warning bits defined here are copied into the Device Block ID 28 in the
"Device Exception Detail Warning" range of the diaphragm gauge measuring part.
3.3.4 Diaphragm Gauge
Transducer Block /
SLOT 1 / Instance 3
3.3.4.1 Block Type
3.3.4.2 Status Extension
3.3.4.3 Sensor Alarm
3.3.4.4 Sensor Warning