INFICON BCG450-SP ATM to Ultra-High Vacuum Triple Gauge (Profibus) User Manual
Page 17

tira41e1 (2005-06) BCG450SPv1.cp
17
Bit
Device Exception Detail 0
Referring to Capacitance
diaphragm sensor
Bit
Device Exception Detail 1
Referring to Capacitance
diaphragm sensor
0
Diaphragm Failure
0
0
1
0
1
Electronics/sensor error
2
0
2
0
3
0
3
0
4
0
4
0
5
0
5
0
6
0
6
0
7
0
7
0
This byte is a copy of Sensor Alarm
byte 0 of Capacitance diaphragm
sensor Transducer Block.
This byte is a copy of Sensor Alarm
byte 1 of Capacitance diaphragm
sensor Transducer Block.
Bit
Device Exception Detail 2
Referring to Pirani
Bit
Device Exception Detail 3
Referring to Pirani
0
0
0
0
1
0
1
Electronics/sensor error Pirani
2
0
2
0
3
0
3
0
4
0
4
0
5
0
5
0
6
0
6
0
7
0
7
0
This byte is a copy of Sensor Alarm
byte 0 of the Heat Transfer Vacuum
Gauge Transducer Block.
This byte is a copy of Sensor Alarm
byte 1 of the Heat Transfer Vacuum
Gauge Transducer Block.
Bit
Device Exception Detail 4
Referring to BA
Bit
Device Exception Detail 5
Referring to BA
0
0
0
0
1
0
1
Electronics/sensor error BA
2
0
2
0
3
0
3
0
4
0
4
0
5
0
5
0
6
0
6
0
7
0
7
0
This byte is a copy of Sensor Alarm
byte 0 of the Hot Cathode Ion Gauge
Transducer Block.
This byte is a copy of Sensor Alarm
byte 1 of the Hot Cathode Ion Gauge
Transducer Block.
Manufacturer Exception Detail 0 is set to "1",
if there is an internal communication
error in the gauge.
Device Exception
Detail Alarm
Manufacturer Exception
Detail Alarm 0