Service
Code
Service Name
Description of Service
32
h
Set Emission
Control Mode
• Automatic
The Emission is switched on and off by the
Pirani automatically.
• Manual
The emission has to be switched on and off
by the user
Parameter
Data Type
Description
Semantics of Values
Emission
User Mode
State
BOOL
→ below
0
Automatic Mode (default)
1
Manual Mode
Automatic:
The emission is automatically switched ON and OFF.
The user can switch the emission OFF by using the Set Emission
State service.
If the emission is switched off manually, the emission will be automa-
tically switched on again, after the pressure rose above 3.2
×10
-2
mbar
and falls below 2.4
×10
-2
mbar again.
Manual:
The emission has to be switched on and off by the user, using the Set
Emission State service. The Set Emission State service may be
executed only, if the pressure is below 2.4
×10
-2
mbar. If the pressure
is higher, an Object State Conflict error will be returned.
If the emission is switched on and the pressure rises above 3.2
×10
-2
mbar, the
gauge will switch off the emission automatically.
The following behavior with respect to Data Type applies:
The Data Type value will be set automatically based upon the first valid I/O con-
nection established by the device.
If no established I/O connections exist, which include an attribute from this object,
then the Data Type attribute is settable provided that the object is in the Idle State.
The following example demonstrates this behavior:
A device specifies an instance of the S-Analog Sensor object as well as two static
Assembly object instances, both with data attribute components mapped to this
object instance. Assembly object instance ID 1 specifies INT data types and As-
sembly object instance ID 2 specifies REAL data types.
After the device is online, it is configured with an I/O connection to Assembly in-
stance ID 2. When the connection transitions to the Established State, this object
instance attribute Data Type is automatically set with the value for REAL before
any data is communicated to, or from the object instance. Any subsequent attempt
to connect to Assembly instance ID 1 would then be rejected and result in an
INVALID ATTRIBUTE VALUE error with the additional error code indicating the ID
of the offending attribute, which in this case would be the connection path.
2.4.4.3 Emission Control Mode
Set Emission Control Mode
Data Field
Emission User Modes
2.4.5 Behavior
Data Type