INFICON BCG450-SD ATM to Ultra-High Vacuum Triple Gauge (DeviceNet) User Manual
Page 11

tira40e1 (2005-04) BCG450SDv1.cp
11
Attribute ID
Access Rule
NV/
V
Name
DeviceNet Data Type
Description of Attribute
14
Get
V
Exception Detail
Warning
STRUCT of:
A structure of three structures containing
a bit mapped representation of the
warning detail
Common Exception
Detail
STRUCT of:
Size 2
USINT
Number of Common Detail Bytes
Detail 0
ARRAY of:
→ "Semantics" section below.
Detail 1
BYTE
→ "Semantics" section below.
Device Exception
Detail
STRUCT of:
Size 7
USINT
Number of Device Detail Bytes
Detail 0
ARRAY of:
S-Analog Sensor
Class Level Status Extension
Detail 1
ARRAY of:
S-Analog Sensor Instance 3
(Capacitance Diaphragm Gauge)
Sensor Warning Byte 0
Detail 2
BYTE
S-Analog Sensor Instance 3
(Capacitance Diaphragm Gauge)
Sensor Warning Byte 1
Detail 3
BYTE
S-Analog Sensor Instance 1 (Pirani)
Sensor Warning Byte 0
Detail 4
S-Analog Sensor Instance 1 (Pirani)
Sensor Warning Byte 1
Detail 5
S-Analog Sensor Instance 2
(hot cathode) Sensor Warning Byte 0
Detail 6
S-Analog Sensor Instance 2
(hot cathode) Sensor Warning Byte 1
Manufacturer
Exception Detail
STRUCT of:
Size 1
USINT
Number of Manufacturer Detail Bytes
Detail 0
Byte
serial common warning
15
Set
NV
Alarm Enable
BOOL
→ "Semantics" section below.
16
Set
NV
Warning Enable
BOOL
→ "Semantics" section below.
This attribute represents the current state of the device. Its value changes as the
state of the device changes. The following values are defined:
Attribute Value
State
0
Undefined
1
Self Testing
2
Idle
3
Self-Test Exception
4
Executing
5
Abort
6
Critical Fault
In case of a Self-Test Exception restart the device with a reset out of the box
(Identity Object Class 1, Instance 1, Service 5, Service target value: 1).
2.3.2.1 Semantics
Device Status