tira40e1 (2005-04) BCG450SDv1.cp

21

Attribute

ID

Access

Rule

NV/
V

Name

DeviceNet
Data Type

Description of Attribute

Semantics of Values

96

Get

V

Status
Extension

BYTE

Bit mapped byte providing
additional status bits

Bit description:
0

Reading Invalid
(Logical Inversion of
Reading Valid)

1

Overrange Exceeded

2

Underrange Exceeded

99

Get

NV

Subclass

UINT

Identifies the subset of
additional attributes, serv-
ices and behaviors for hot
cathode ion gauges

5

Hot Cathode Ion Gauge

100

Get

NV

State
Emission
Control
Mode

USINT

Represents the state ac-
cording the service Set
Control Mode

0 = Auto

1 = Manual

→ Instance 1

An S-Analog Sensor object instance derives a reading from a physical analog sen-
sor. The reading is converted to the data type and units specified for the Value
attribute.

Using Counts and INT the following conversion has to be used:

Counts

=

[ log

10

(pressure) + k ] × 2000

where:

k

mbar

=

12.5

k

Torr

=

12.624903

k

Pa

=

10.5

→ Instance 1

→ Instance 1

16 Bits are used as sensor faults . Sensor Alarm Byte 0 is mapped to the Exception
Detail Alarm 4, Sensor Alarm Byte 1 ismapped to the Exception Detail Alarm 5.

Data Component

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

Sensor Alarm Byte 0

0

0

0

0

0

0

0

0

Sensor Alarm Byte 1

0

0

0

0

0

0

Electronics

Failure

0

16 Bits are used as sensor warnings. Sensor warning Byte 0 is mapped to the
Exception Detail Warning 5, Sensor warning Byte 0 is mapped to the Exception
Detail Warning 6.

Data Component

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

Sensor Warning Byte 0

0

0

0

0

0

0

0

0

Sensor Warning Byte 1

0

0

0

0

Pressure
too high for
degas

0

0

0

The bit "Pressure too high for degas" will be set if the pressure is above
7.2×10

-6

mbar when a degas service is requested. The bit will be reset when the

pressure is below 7.2×10

-6

mbar.

2.4.2.4 Semantics of S-Analog

Sensor Instance 2

Data Type

Value

Safe State

Safe Value

Sensor Alarm

Sensor Warning