INFICON BCG450-SD ATM to Ultra-High Vacuum Triple Gauge (DeviceNet) User Manual
Page 21
tira40e1 (2005-04) BCG450SDv1.cp
21
Attribute
ID
Access
Rule
NV/
V
Name
DeviceNet
Data Type
Description of Attribute
Semantics of Values
96
Get
V
Status
Extension
BYTE
Bit mapped byte providing
additional status bits
Bit description:
0
Reading Invalid
(Logical Inversion of
Reading Valid)
1
Overrange Exceeded
2
Underrange Exceeded
99
Get
NV
Subclass
UINT
Identifies the subset of
additional attributes, serv-
ices and behaviors for hot
cathode ion gauges
5
Hot Cathode Ion Gauge
100
Get
NV
State
Emission
Control
Mode
USINT
Represents the state ac-
cording the service Set
Control Mode
0 = Auto
1 = Manual
→ Instance 1
An S-Analog Sensor object instance derives a reading from a physical analog sen-
sor. The reading is converted to the data type and units specified for the Value
attribute.
Using Counts and INT the following conversion has to be used:
Counts
=
[ log
10
(pressure) + k ] × 2000
where:
k
mbar
=
12.5
k
Torr
=
12.624903
k
Pa
=
10.5
→ Instance 1
→ Instance 1
16 Bits are used as sensor faults . Sensor Alarm Byte 0 is mapped to the Exception
Detail Alarm 4, Sensor Alarm Byte 1 ismapped to the Exception Detail Alarm 5.
Data Component
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Sensor Alarm Byte 0
0
0
0
0
0
0
0
0
Sensor Alarm Byte 1
0
0
0
0
0
0
Electronics
Failure
0
16 Bits are used as sensor warnings. Sensor warning Byte 0 is mapped to the
Exception Detail Warning 5, Sensor warning Byte 0 is mapped to the Exception
Detail Warning 6.
Data Component
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Sensor Warning Byte 0
0
0
0
0
0
0
0
0
Sensor Warning Byte 1
0
0
0
0
Pressure
too high for
degas
0
0
0
The bit "Pressure too high for degas" will be set if the pressure is above
7.2×10
-6
mbar when a degas service is requested. The bit will be reset when the
pressure is below 7.2×10
-6
mbar.
2.4.2.4 Semantics of S-Analog
Sensor Instance 2
Data Type
Value
Safe State
Safe Value
Sensor Alarm
Sensor Warning