Service
Code
Service Name
Description of Service
32
h
User Atmosphere
Adjust
Performs an Atmosphere Adjust Piezo to
Capacitance Diaphragm Gauge
User ATM Adjust Service Data Field Parameters:
Parameter
Datatype
User ATM Adjust
BOOL
0 = Unlock user ATM Adjust
1 = executes an User ATM Adjust and lock
There are no state transitions associated with the invocation of this service. It is
therefore, incumbent upon the user to establish the device into the desired con-
figuration prior to, and during the execution of this service. This will generally in-
volve exposing the sensor to a known environment and treating the values read
during execution of the services accordingly.
A success service response indicates that the service was accepted and the ap-
plication process started.
To perform the User Atmosphere Adjust Request proceed as follows:
n
Vent the gauge to atmosphere
o
Unlock the user ATM Adjust wit a target value of "0"
p
Execute the User Atmosphere Adjust Service with a target value of "1".
The pressure value of the amtmosphere pressure sensor is then tracked to the
value of the Capacitance Diaphragm Gauge.
2.4.2.7 Object-Specific Services
on Instance 3 / Capacitance
Diaphragm Gauge
USER ATM Adjust