INFICON Crystal 12 Sensor User Manual
Page 48
4 - 4
07
4-
39
8-
P1A
Crystal12 Sensor Operating Manual
6
Actuate the bellows assembly by applying 24 volts to the leads of the solenoid
valve of the solenoid valve assembly. Consult
for wiring
assignments. The air pressure must be sustained.
7
Rotate the ratchet (item 29,
) counterclockwise until one
tooth contacts the pin of the pawl and actuator stem assembly (item 18,
). Secure the ratchet to the 0.25" diameter shaft of the carousel
spindle body (item 24,
) utilizing the #6-32 x 0.125" long
).
8
Rotate the stop ratchet (item 31,
) clockwise until it
contacts the stem of the pawl and actuator stem assembly (item 18,
) as shown in
. Secure the stop ratchet to the 0.25”
diameter shaft of the carousel spindle body (item 24,
)
utilizing the #6-32 x 0.125 long socket set screws (item 12,
).
Figure 4-2 Setting the Stop Ratchet
9
Position the detent spacing tool (of the 750-254-G2 assembly) between the
detent (item 25,
) and ratchet (item 29,
) as shown in
. Pull the detent against the detent spacing
tool and subsequently against the tooth of the ratchet. Tighten the #2 fastener
(item 5,
) to secure the detent to the bellow assembly.