2 procedure – INFICON Crystal 12 Sensor User Manual
Page 47
4 - 3
07
4-
39
8-
P1A
Crystal12 Sensor Operating Manual
Alignment tool assembly 750-254-G2 (Included in 750-268 kit)
Regulated air supply 80-90 PSIG (5.5 bar - 6.2 bar) [550 kPa - 620 kPa]
NOTE: 75 PSIG (5.2 bar) [520 kPa] is for alignment purposes only.
4.3.2 Procedure
With the required equipment at hand, proceed as follows:
1
Temporarily connect an air supply to the bellows assembly (item 37,
) supply tube. Regulate the air pressure to 75 PSIG (5.2 bar) [520
kPa].
2
Insert the 0.09375" diameter alignment pin (of the 750-254-G2 assembly) in the
alignment hole (if not already installed) as shown in
. The crystal
carousel may need to be rotated to achieve the alignment condition shown in
Figure 4-1 Alignment Pin Installation
3
Remove the actuator cover (item 22,
) and its related
hardware (items 6 and 7,
4
Loosen the set screws (two set screws per item) of the ratchet (item 29,
) and the stop ratchet (item 31,
5
Loosen the #2 fastener (item 5,
) that attaches the detent
(item 25,
) to the bellows assembly item 37,
). Do NOT remove detent.
Crystal Carousel
0.09375 dia. Alignment Pin
750-254-G2
Carousel
Housing
Weldment
Front
Deposition
Shield