T3000, T3000 configurations – INFICON T3000 Thermal Power Supply User Manual
Page 40

instruments
Thermal Power Supply
T3000
Section 7
Page 7 - 1
T3000 Configurations
Section 7
T3000 Configurations
This section provides a procedure for operation of the T3000 in various configurations.
Please refer to the system diagram in Section 2.1-1.
7.1 T3000
Stand-Alone Configuration
1) Verify that the AC input power is disabled.
2) Connect the AC-In power cable to the T3000 as shown in Figure 2.2-1.
3) Connect the output buss bars for the desired configuration, series or parallel, as per
Figures 2.2-3 or 2.2-4.
4) Connect the control cable to J1 as shown in Figure 3.2-1. (Note: Voltages measured on
pins 5 and 14 are used as monitor points.)
5) Apply the AC power. The green LED should now “blink” indicating that AC power is
applied to the T3000.
6) With 0 V applied to pin 7 of J1, close the Off/On switch. The green LED should now be
on in a steady state which indicates that the T3000 is ready to supply current to the load.
7) Gradually apply the desired input voltage, 0 to +10 V, to pin 7 of J1. (Note: For the
series mode, 1 volt in = 50 ARMS out while 1 volt in = 100 ARMS out for the parallel mode.)
8) When the operation is complete, turn off current to the load by opening the “On/Off”
switch and apply 0 volts to pin 7 of J1. (Note: The operation can also be terminated by
applying 0 volts to pin 7 and then opening the “On/Off” switch.