Installation
Page 2 - 2
Section 2
Table 2.2-2
T3000 Connection Interface
Connector Function
Description
AC line connections & frame ground
Output Bus Bars
J1 control interface
8-32 Threaded Stand-off accepts #8
Ring terminal
¼” bolt holes
15 pin DB female (accepts 15 pin male
matching connector on cable)
Figure 2.2-1 shows the location and connection configuration for the single AC line input
(180-264 VAC). The AC wiring should be terminated with NO. 8 ring terminals and secured
to the 8-32 male terminals on the T3000 with nut and lock-washer. A safety cover, with
strain relief, covers the AC input connections.
CAUTION
Figure 2.2-2 shows the location of the bus bars for the two output sections and defines the
polarity relative to each other. Note that the Output-2 positive bus bar is common with the
T3000 signal ground.
CAUTION
Figure 2.2-3 shows how the auxiliary C-shaped and flat bus bars are utilized to connect the
two output sections of the T3000 in the parallel configuration. Cables to the thermal
sources (usually the feedthrough terminals of the vacuum chamber) must be terminated in
a suitably current rated solderless connector with a ¼-inch bolt hole for connection to the
bus bars.
Hazardous
Voltages
Present
Hazardous
Current
Levels
Possible
TURN OFF THE INPUT AC POWER WITH THE
SAFETY SWITCH BEFORE REMOVING THE
T3000 PROTECTIVE COVER.
TURN OFF THE INPUT AC POWER
WITH THE SAFETY SWITCH
BEFORE CHANGING
CONNECTIONS TO BUS BARS.