beautypg.com

3 changing probe protection cap, Notice, Changing probe protection cap – INFICON Sensistor Sentrac Hydrogen Leak Detector User Manual

Page 62

background image

60

ni

n65

e1-a (140

9)

14.3

Changing Probe Protection Cap

Disassemble

1

Attach the supplied Sensor Key at the bottom of the Probe Tip Protection Cap (A)
and pull with one finger on each side of the probe tip.

2

Remove the Protection Cap (A).

Assemble

1

Install the new Probe Tip Protection Cap by pressing it over the tip until you hear a
clicking sound.

Part

Part no.

Qty

Consumables

Probe Tip Protection Cap

591-273 (set of 50)

1

590-625 (set of 500)

1

A

A

Notice

If the probe tip filter is removed, an air gap is formed between the inner
metal filter and filter in the protection cap.

The filter in the protection cap prevents water to get into the probe if the
probe comes in contact with water.

It is recommended to always replace the protection cap with a new if
exposed to water.