2 hardware error messages, Hardware error messages – INFICON Sensistor Sentrac Hydrogen Leak Detector User Manual
Page 58
56
ni
n65
e1-a (140
9)
13.2
Hardware Error Messages
Warning
Message
Indicators
Message in the menu
warnings list
Error Code
Alert
Probe not connected.
Blue LED flashing.
(None)
(None)
Alert
SD-card not inserted.
Orange measure
window at screen.
(None)
(None)
Alert
Battery discharged. Recharge
battery.
Orange measure
window at screen, fast
Blue LED flashing.
(None)
(None)
Alert
Low sensor signal.
Reject level to low for
sensor sensitivity.
(None)
Alarm
Sensor broken or not connected.
Change sensor or connect
sensor. (Voltage to high, Error
E1)
Red measure window
at screen, Blue LED
flashing.
Sensor broken or not
connected. (Voltage to
high)
01
Alarm
Sensor broken. Change sensor.
(Voltage to Low, Error E2)
Red measure window
at screen, Blue LED
flashing.
Sensor broken.
(Voltage to Low)
02
Alarm
Sensor broken. Change sensor.
(Temperature to Low, Error E3)
Red measure window
at screen, Blue LED
flashing.
Sensor broken.
(Temperature error)
03
Alarm
Probe disconnected. Reconnect
the probe. (Error 04)
Red measure window
at screen, Blue LED
flashing.
Probe disconnected.
04
Alarm
Probe communication problem.
Try restart instrument or change
cable. If the problem remains
contact service. (Error 05)
Red measure window
at screen, Blue LED
flashing.
Probe communication
problem.
05
Alarm
Probe Software not compatible
with instrument. Change probe
or upgrade probe software.
(Error 06)
Red measure window
at screen, Blue LED
flashing.
Probe Software not
compatible with
instrument.
06
Alarms
Two probes connected.
Disconnect probe and restart
instrument. (Error 07)
Red measure window
at screen, Blue LED
flashing.
Two probes connected. 07
Alarms
Instrument hardware error,
service recommended (Error 08)
Red measure window
at screen, Blue LED
flashing.
Instrument hardware
error.
08
Alarms
Probe hardware error, service
recommended. (Error 09)
Red measure window
at screen, Blue LED
flashing.
Probe hardware error.
09