11 info, 1 statistics, 1 run time – INFICON Sensistor Sentrac Hydrogen Leak Detector User Manual
Page 53: 2 calibration, 3 i·guide, 2 export/import, 1 export, 2 import, Info, Statistics
51
ni
n65
e1-a (140
9)
11
Info
11.1
Statistics
11.1.1
Run Time
To see the current operation time, click Info >> Statistics >> Run Time.
Two runtime counters can be reset and used for specific needs. The two timers are
also reset when a reset is performed on the instrument.
11.1.2
Calibration
To see information regarding:
•
number of calibrations in total
•
number of failed calibrations
•
date and time of latest calibration
•
etc.
Click Info >> Statistics >> Calibration.
11.1.3
I·Guide
To see statistics about:
•
number of tested objects
•
number of accepted tests
•
number of rejected tests
Click Info >> Statistics >> I·Guide.
11.2
Export/Import
11.2.1
Export
Backup of all useful settings, including recipes, can be exported to an SD-card.
1
Insert an SD-card in the SD-card holder
2
Click Info >> Export/Import >> Export
3
In Settings, click on the backup to SD-button
A text file is created on the SD-card.
The settings can be imported to any Sentrac instrument.
11.2.2
Import
All useful settings, including recipes, can be imported from an SD-card.
1
Insert the SD-card with the backup into the SD-card holder
2
Click Info >> Export/Import >> Import