Installation – Det-Tronics MOS H2S Gas Sensor User Manual

Page 6

background image

95-8532

1.4

4

Two additional options are available for resetting the

transmitter. A user supplied switch can be wired to the

transmitter terminal block and used to reset the device

from a remote location. In addition, Detector Electronics

offers a special optional enclosure with a pushbutton

pre-installed within a multi-port electrical enclosure for

customers who do not prefer magnetic switches.

The U9500B utilizes a serial EEPROM for storing

configuration data, setpoints, and calibration data.

MODES Of OpERaTION
The U9500B has three main operating modes:

- Normal

- Calibrate

- Setup (also called Configuration).

Normal mode is the default mode (no buttons pushed).

Three minor modes also exist:

- Setpoint display

- Reset

- Power up.

INSTALLATION

WIRING REQUIREMENTS
• Install per local installation practices and in

accordance with local authority having jurisdiction.

• See Figure 2 for wiring size and maximum distance

from the power supply to the transmitter.

IMPORTANT

Proper installation practices must be followed
to ensure that condensation does not enter the
junction box and interfere with the electronic
circuitry. Refer to local installation codes.

WIRING pROCEDURE
The following procedure should be used for installing

and wiring the Infiniti Transmitter.

NOTE

The U9500B with MOS sensor does not support
sensor separation i.e. the sensor must be attached
directly to the sensor housing.

Install the unit in a location that is best suited to cover

1.

the area to be protected. Whenever practical, the

Infiniti should be placed where it is easily accessible

for calibration. For proper operation, sensors

should be pointing down. See Figure 1 for mounting

dimensions.
Remove the transmitter from its packing material and

2.

note the location of the field wiring terminals.
Remove the cover from the enclosure. One conduit

3.

entry is used for mounting the sensor and the other

is for field wiring. The transmitter module slides into

the mounting bracket within the enclosure and all

field and sensor wiring connects to this module. The

sensor connects to a small plug on one side of the

transmitter and all field wiring connects to terminals

on the other side.

The transmitter can be oriented either way within

the bracket (180° difference) and the bracket can

be rotated 90° to achieve the necessary orientation.

See Figure 3 for left and right field wiring entry and

corresponding bracket orientation. Make sure that

the display is oriented correctly, then proceed with

transmitter wiring.

Mount the transmitter enclosure and connect the field

4.

wiring conduit. The enclosure should be electrically

connected to earth ground.

0

DISTANCE (FEET)

2000

4000

6000

8000

10000

12 AWG

14 AWG

16 AWG

18 AWG

POWER SUPPLY VOLTAGE

32

30

28

26

24

22

20

18

16

14

A1672

Figure 2—Wiring Requirements for Transmitters with Relays