beautypg.com

Det-Tronics MOS H2S Gas Sensor User Manual

Page 20

background image

95-8532

1.4

18

CALIBRATION

fREQUENCY Of CaLIbRaTION
The MOS H

2

S sensor is typically utilized to protect human

life. For this reason, a frequent calibration inspection

is recommended. The specific frequency required in

different applications can vary depending upon the

amount of background gas, concentration of exposed

H

2

S, and ambient environmental conditions.

Calibration must be performed:

When a new system is initially put into service.

When the sensor is replaced.

Periodically to verify proper performance.

To ensure optimum performance, the recommended

frequency of calibration for MOS H

2

S sensors is:

After a one hour warm-up period.

After 24 hours.

After one week.

Every 90 days.

IMPORTANT

To ensure adequate protection, the H

2

S detection

system must be calibrated on a regularly
scheduled basis.

CaLIbRaTION GaS
The U9500B/MOS H

2

S sensor must be calibrated using

only Det-Tronics 40 ppm H

2

S ampoules. All ampoule

calibrations must be performed using Det-Tronics

calibration bottle with thumb screw ampoule breaker and

internal mixing fan. Do not use bottled 40 ppm H

2

S in

nitrogen.

aUTO CaLIbRaTION pROCEDURE
The Infiniti with MOS H

2

S sensor is calibrated using the

Automatic Calibration sequence. All adjustments are

made automatically by the transmitter once calibration

has been initiated. Refer to Table 7 for the Automatic

Calibration Procedure.

IMpORTaNT CaLIbRaTION NOTES

Ensure that only clean air is present at the sensor prior

to entering the calibration mode. If the possibility

of background gases exists, purge the sensor with

clean air to ensure accurate calibration.
If the calibration procedure is not completed or

if the sensitivity of the sensor has deteriorated to

the extent that calibration cannot be successfully

completed, a fault will be generated and the system

will automatically revert back to the former calibration

settings (after 10 minutes or when the gas level drops

below the lowest setpoint). If a successful calibration

cannot be accomplished, replace the sensor and re-

calibrate.
Inspect the sensor. Loss of sensitivity can be

caused by various factors. One common cause is

by clogging of the hydrophobic or sintered filter by

dirt, oil, paint, etc. Problems of this nature will not

be detected by the transmitter’s diagnostic circuitry.

If the filter or the sensor are dirty or plugged, they

should be replaced.