beautypg.com

Det-Tronics PIR9400 PointWatch Infrared Hydrocarbon Gas Detector User Manual

Page 25

background image

23

95-8440

9.3

Current Level

Status

Corrective action

2.4 to 3.9 mA

Negative Zero Indication

NOTE:

This phenomenon is typically caused either by the presence

of background gas during zero calibration, or by condensation on the

device optics. If low level background hydrocarbon gas was present
during calibration, the result will be a signal output level below

4 mA

when the background gas clears. To correct this, the device must be

re-calibrated with all background gas removed. Purge the detector
optics with Det-Tronics “zero air” calibration gas for about

30 seconds

prior to initiating calibration.

In outdoor environments having high humidity coupled with rapid

temperature swings, very small amounts of condensation may form on

the optics, causing a temporary (up to a few hours) negative excursion
below

4 mA. This phenomenon typically does not cause a loss of

detection capability, and will self-correct as the heated optics dry out
the residual condensation. Excursions down to the

3.0 mA level may

occur without significant loss of detection capability; therefore, it is

recommended that “zero drift” alert messages should be set no higher
than

3.0 mA, with typical threshold settings between 2.4 and 3.0 mA.

The PIR

9400 hydrophobic filter offers significant protection against

condensation. Verify that the hydrophobic filter is properly installed

and not crumpled or distorted to prevent a direct humidity path into the

device optics.

1.6 mA

Calibration fault

Make sure that the calibration gas being used matches the Gas

Selection Switch setting. If these match and the fault is still present,

perform disassembly and cleaning procedure, then recalibrate.

1.0 mA

Fouled optics

Perform disassembly and cleaning procedure, then recalibrate.

0.8 mA

+24 Vdc line low (less than +17.5 Vdc)

Ensure that input voltage is correct and that power connections are

good. If fault does not clear, replace the electronics assembly.

0.6 mA

Calibrate input active at power-up

Ensure that calibration line is not shorted and that the calibration switch

is open. If fault does not clear, replace the unit.

0.4 mA

Active channel fault

Replace electronics assembly.

0.2 mA

Reference channel fault

Replace electronics assembly.

0.0 mA

CPU system fault, warmup

Ensure that power is applied and that the warmup period is complete
(

1 minute). If fault does not clear, replace the unit.

Table 5—Troubleshooting Table