beautypg.com

Dakota Ultrasonics DFX-7 plus MANUAL2 User Manual

Page 75

background image

DFX-7

Ultrasonic Thickness Gauge

71

5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.

6) Repeat steps 4 & 5 until the DELAY value is correctly displayed.

7) Press the OK key to return to the measurement screen, or ESC to cancel

entering the DELAY.

8) Finally, press the MEAS key to return to the measurement screen and begin

taking readings.


Adjusting the Range (B-Depth)

Once again, just to reiterate, the Range (B-Depth) + Delay (B-Start) equals the
right side of the screen
. Therefore, the Range (B-Depth) is the overall area, from
the delay, that will be viewable on the screen. The DFX-7 digitizer will round up from
the Range (B-Depth) that’s entered. Therefore, if the Range (B-Depth) is set at 1.0”,
the digitizer will round this value up to the next adjustment available. The procedures
to adjust the overall thickness range viewed Range (B-Depth) are outlined below:

Adjusting the Range (B-Depth) using the Hot Menus

1) Press the MEAS key once to activate measure menu items. Press the

MEAS key multiple times to move right and the ESC key multiple times to

move left, until the WIDTH cell is highlighted.

This manual is related to the following products: